Description
Leybold Vacuum TMP 361 and TMP 151 Manual
Leybold Vacuum TURBOVAC 151 (C) Technical Data and Ordering Information
Leybold Vacuum TURBOVAC Classic Line Product Information
Leybold Vacuum TURBOVAC Line Operating Instructions
Technical Data:
Technical Data | TURBOVAC 151 (C) | |
---|---|---|
Connection Inlet DN | 100 ISO-K | 100 CF |
Connection Outlet DN | 25 ISO-KF | 25 ISO-KF |
Pumping speed N₂ | 145 l/s | 145 l/s |
Pumping speed Ar | 150 l/s | 150 l/s |
Pumping speed He | 135 l/s | 135 l/s |
Pumping speed H₂ | 115 l/s | 115 l/s |
Gas throughput N₂ | 1.5 mbar l/s | 1.5 mbar l/s |
Gas throughput Ar | 1.3 mbar l/s | 1.3 mbar l/s |
Gas throughput He | 1.5 mbar l/s | 1.5 mbar l/s |
Gas throughput H₂ | 1.0 mbar l/s | 1.0 mbar l/s |
Compression ratio N₂ | 1 x 10⁹ | 1 x 10⁹ |
Compression ratio Ar | 1 x 10⁹ | 1 x 10⁹ |
Compression ratio He | 2 x 10⁴ | 2 x 10⁴ |
Compression ratio H₂ | 8 x 10² | 8 x 10² |
Ultimate pressure | < 1.0 x 10⁻¹⁰ mbar (<7.5 x 10⁻¹¹ Torr) | < 1.0 x 10⁻¹⁰ mbar (<7.5 x 10⁻¹¹ Torr) |
Max. foreline pressure for N₂ | 5.0 x 10⁻¹ mbar (3.8 x 10⁻¹ Torr) | 5.0 x 10⁻¹ mbar (3.8 x 10⁻¹ Torr) |
Recommended foreline pump | from TRIVAC D 4 B to D 16 B | from TRIVAC D 4 B to D 16 B |
Nominal rotation speed | 50000 rpm | 50000 rpm |
Run-up time | 2 min | 2 min |
Max. power consumption | 300 W | 300 W |
Power consumption at ultimate pressure | 70 W | 70 W |
Admissible ambient temperature | 10 to 55 °C / 50 to 131 °F | 10 to 55 °C / 50 to 131 °F |
Cooling standard | Water | Water |
Cooling optional | Air | Air |
Cooling water connection | 10 mm hose nozzle | 10 mm hose nozzle |
Cooling water consumption | 15 to 35 l/hr | 15 to 35 l/hr |
Permissible cooling water pressure | 3 to 7 bar | 3 to 7 bar |
Permissible cooling water temperature | 10 to 25 °C (50 to 77 °F) | 10 to 25 °C (50 to 77 °F) |
Weight | 8 kg (17 lbs) | 8 kg (17 lbs) |