Products

Leybold TURBOVAC 1000C

The Leybold TURBOVAC 1000 C turbomolecular pump can be used for all general-purpose high vacuum applications or for pumping aggressive and reactive gases. This model comes with a purge gas facility for use with aggressive gases. It can produce an ultimate pressure below 10^-10 mbar. It can achieve a pumping speed of 1000 liters/second for air at molecular flow.  

Services Offered:

  • Sale of refurbished product
  • Repair Service

Description

We offer refurbished TURBOVAC 1000 C pumps in most configurations. Contact us today for current price and lead times. We have several refurbished models available for $9,500.

Features:

  • High pumping speed and throughput
  • Installation in any orientation
  • Features grease lubricated ceramic bearings
  • Easy to operate and very reliable
  • Robust rotor design

Applications:

  • Data storage
  • Flat panel displays
  • Optical coating
  • Large Area Coating
  • Fusion experiments
  • Space simulation
  • Load locks

Common configurations we offer include:

  • ISO 160 housing with KF40 foreline (PN 85535)
  • ISO 160 housing with ISO63 foreline (PN 85538)
  • ISO 200 housing with KF40 foreline (PN 15300)
  • ISO 250 housing with KF40 foreline (PN 85536)
  • ISO 250 housing with ISO63 foreline (PN 85539)
  • CF 160 housing with KF40 foreline (PN 85491)
  • CF 200 housing with KF40 foreline (PN 11764)
  • 6in ASA housing with KF40 foreline (PN 89489)
  • 6in ASA housing with ISO63 foreline (PN 89499)

Our TURBOVAC rebuild/repair service includes:

  • Complete cleaning of the pump
  • Replacement of all O-rings, seals, and ceramic bearings
  • The pump is balanced to meet or exceed Leybold vibration specifications
  • The pump is run overnight on test
  • Leak tested to 1 x 10^-10 mbar l/s
  • Final vibration data sheet provided with each pump
  • Every pump is refurbished to like new condition
  • We stock new OEM Leybold rotors, stator blades, stator rings, motors, shafts, motor rotors, water lines and more. We have every part needed to refurbish your pump to like new no matter what condition your pump is currently in.

Manuals:

Leybold Vacuum TMP 1000 Turbopump Manual

Leybold Vacuum TURBOVAC 1000 C Technical Data and Ordering Information

Leybold Vacuum TURBOVAC Classic Line Product Information

Leybold Vacuum TURBOVAC Line Operating Instructions

Technical Data:

Technical Data TURBOVAC 1000 C
Connection

      Inlet                                 DN

 

     Outlet                               DN

 

 

160 ISO-K ● 160 CF

 

40 ISO-KF ● 

63 ISO-K

 

 

200 ISO-K ● 200 CF

 

40 ISO-KF ●

 63 ISO-K

 

 

250 ISO-K

 

40 ISO-KF ●

 63 ISO-K

Pumping speed

      N2                    l x s^-1

      Ar                     l x s^-1

      He                    l x s^-1

      H                  l x s^-1

 

850

810

880

900

 

1100

1050

975

970

 

1150

1100

1000

1000

Gas throughput

      N2                   mbar x l x s^-1

      Ar                    mbar x l x s^-1

      He                   mbar x l x s^-1

      H                mbar x l x s^-1

 

6.5

4.0

7.0

8.0

 

6.5

4.0

7.0

8.0

 

6.5

4.0

7.0

8.0

Compression ratio

      N2                      

      Ar                     

      He                    

      H                   

 

> 1 x 10^9

> 1 x 10^9

5 x 10^4

1 x 10^4

 

> 1 x 10^9

> 1 x 10^9

5 x 10^4

1 x 10^4

 

> 1 x 10^9

> 1 x 10^9

5 x 10^4

1 x 10^4

Ultimate pressure    mbar (Torr) < 1 x 10^-10 (<7.5 x 10^-11) < 1 x 10^-10 (<7.5 x 10^-11) < 1 x 10^-10 (<7.5 x 10^-11)
Max. foreline pressure for N2 

                                 mbar (Torr)

5 x 10^-2 (3.8 x 10^-2) 5 x 10^-2 (3.8 x 10^-2) 5 x 10^-2 (3.8 x 10^-2)
Recommended foreline pump from TRIVAC D 25 B to D 65 B from TRIVAC D 25 B to D 65 B from TRIVAC D 25 B to D 65 B
Nominal rotation speed        rpm 36000 36000 36000
Run-up time                         min 9 9 9
Max. power consumption       W 300 300 300
Power consumption at ultimate pressure                                 W 200 200 200
Admissible ambient temperature

                                         °C(°F)

10 to 55 (50 to 131) 10 to 55 (50 to 131) 10 to 55 (50 to 131)
Cooling

      standard

      optional

 

Water

Air

 

Water

Air

 

Water

Air

Cooling water connection 10 mm hose nozzle 10 mm hose nozzle 10 mm hose nozzle
Cooling water consumption  l/hr 20 to 80 20 to 80 20 to 80
Permissible cooling water pressure                                bar 3 to 7 3 to 7 3 to 7
Permissible cooling water temperature                     °C(°F) 10 to 30 (50 to 86) 10 to 30 (50 to 86) 10 to 30 (50 to 86)