Description
Leybold Vacuum TURBOVAC 50 Technical Data and Ordering Information
Leybold Vacuum TURBOVAC Classic Line Product Information
Leybold Vacuum TURBOVAC Line Operating Instructions
Technical Data | TURBOVAC 50 | |
---|---|---|
Connection Inlet DN | 40 ISO-K / 40 CF | 63 ISO-K / 63 ICF |
Connection Outlet DN | 16 ISO-KF | 16 ISO-KF |
Pumping speed N₂ | 33 l/s | 55 l/s |
Pumping speed Ar | 30 l/s | 50 l/s |
Pumping speed He | 36 l/s | 48 l/s |
Pumping speed H₂ | 28 l/s | 30 l/s |
Gas throughput N₂ | 0.90 mbar l/s | 1.00 mbar l/s |
Gas throughput Ar | 0.70 mbar l/s | 0.80 mbar l/s |
Gas throughput He | 0.30 mbar l/s | 0.40 mbar l/s |
Gas throughput H₂ | 0.25 mbar l/s | 0.30 mbar l/s |
Compression ratio N₂ | 2 x 10⁶ | 2 x 10⁶ |
Compression ratio Ar | 2 x 10⁶ | 2 x 10⁶ |
Compression ratio He | 5 x 10² | 5 x 10² |
Compression ratio H₂ | 2 x 10² | 2 x 10² |
Ultimate pressure | < 5 x 10⁻⁸ mbar (<3.8 x 10⁻⁸ Torr) | < 5 x 10⁻⁸ mbar (<3.8 x 10⁻⁸ Torr) |
Max. foreline pressure for N₂ | 1 x 10⁻¹ mbar (7.5 x 10⁻² Torr) | 1 x 10⁻¹ mbar (7.5 x 10⁻² Torr) |
Recommended foreline pump | TRIVAC D 2.5 E | TRIVAC D 2.5 E |
Nominal rotation speed | 72000 rpm | 72000 rpm |
Run-up time | 2 min | 2 min |
Max. power consumption | 45 W | 45 W |
Power consumption at ultimate pressure | 15 W | 15 W |
Admissible ambient temperature | 10 to 55 °C / 50 to 131 °F | 10 to 55 °C / 50 to 131 °F |
Cooling standard | Convection | Convection |
Cooling optional | Air / Water | Air / Water |
Cooling water connection | 10 mm hose nozzle | 10 mm hose nozzle |
Cooling water consumption | 15 to 25 l/hr | 15 to 25 l/hr |
Permissible cooling water pressure | 3 to 7 bar | 3 to 7 bar |
Permissible cooling water temperature | 10 to 35 °C (50 to 95 °F) | 10 to 35 °C (50 to 95 °F) |
Weight | 2.0 kg (4.4 lbs) | 2.0 kg (4.4 lbs) |