Products

Leybold TURBOVAC 1000C

The Leybold TURBOVAC 1000 C turbomolecular pump can be used for all general-purpose high vacuum applications or for pumping aggressive and reactive gases. This model comes with a purge gas facility for use with aggressive gases. It can produce an ultimate pressure below 10^-10 mbar. It can achieve a pumping speed of 1000 liters/second for air at molecular flow.  

Services Offered:

  • Sale of refurbished product
  • Repair Service

Description

We offer refurbished TURBOVAC 1000 C pumps in most configurations. Contact us today for current price and lead times.

Features:

  • High pumping speed and throughput
  • Installation in any orientation
  • Features grease lubricated ceramic bearings
  • Easy to operate and very reliable
  • Robust rotor design

Applications:

  • Data storage
  • Flat panel displays
  • Optical coating
  • Large Area Coating
  • Fusion experiments
  • Space simulation
  • Load locks
  • Research systems

Common configurations we offer include:

  • ISO 160 housing with KF40 foreline (PN 85535)
  • ISO 160 housing with ISO63 foreline (PN 85538)
  • ISO 200 housing with KF40 foreline (PN 15300)
  • ISO 250 housing with KF40 foreline (PN 85536)
  • ISO 250 housing with ISO63 foreline (PN 85539)
  • CF 160 housing with KF40 foreline (PN 85491)
  • CF 200 housing with KF40 foreline (PN 11764)
  • 6in ASA housing with KF40 foreline (PN 89489)
  • 6in ASA housing with ISO63 foreline (PN 89499)

Our TURBOVAC rebuild/repair service includes:

  • Complete cleaning of the pump
  • Replacement of all O-rings, seals, and ceramic bearings
  • The pump is balanced to meet or exceed Leybold vibration specifications
  • The pump is run overnight on test
  • Leak tested to 1 x 10^-10 mbar l/s
  • Final vibration data sheet provided with each pump
  • Every pump is refurbished to like new condition
  • We stock new OEM Leybold rotors, stator blades, stator rings, motors, shafts, motor rotors, water lines and more. We have every part needed to refurbish your pump to like new no matter what condition your pump is currently in.

Technical Data:

Technical Data TURBOVAC 1000 C
Connection Inlet DN 160 ISO-K / 160 CF 200 ISO-K / 200 CF 250 ISO-K
Connection Outlet DN 40 ISO-KF / 63 ISO-K 40 ISO-KF / 63 ISO-K 40 ISO-KF / 63 ISO-K
Pumping speed N₂ 850 l/s 1100 l/s 1150 l/s
Pumping speed Ar 810 l/s 1050 l/s 1100 l/s
Pumping speed He 880 l/s 975 l/s 1000 l/s
Pumping speed H₂ 900 l/s 970 l/s 1000 l/s
Gas throughput N₂ 6.5 mbar l/s 6.5 mbar l/s 6.5 mbar l/s
Gas throughput Ar 4.0 mbar l/s 4.0 mbar l/s 4.0 mbar l/s
Gas throughput He 7.0 mbar l/s 7.0 mbar l/s 7.0 mbar l/s
Gas throughput H₂ 8.0 mbar l/s 8.0 mbar l/s 8.0 mbar l/s
Compression ratio N₂ > 1 x 109 > 1 x 109 > 1 x 109
Compression ratio Ar > 1 x 109 > 1 x 109 > 1 x 109
Compression ratio He 5 x 10⁴ 5 x 10⁴ 5 x 10⁴
Compression ratio H₂ 1 x 104 1 x 104 1 x 104
Ultimate pressure < 1.0 x 10⁻¹⁰ mbar (<7.5 x 10⁻11 Torr) < 1.0 x 10⁻¹⁰ mbar (<7.5 x 10⁻11 Torr) < 1.0 x 10⁻¹⁰ mbar (<7.5 x 10⁻11 Torr)
Max. foreline pressure for N₂ 5 x 10⁻2 mbar (3.8 x 10⁻² Torr) 5 x 10⁻2 mbar (3.8 x 10⁻² Torr) 5 x 10⁻2 mbar (3.8 x 10⁻² Torr)
Recommended foreline pump from TRIVAC D 25 B to D 65 B from TRIVAC D 25 B to D 65 B from TRIVAC D 25 B to D 65 B
Nominal rotation speed 36000 rpm 36000 rpm 36000 rpm
Run-up time 9 min 9 min 9 min
Max. power consumption 300 W 300 W 300 W
Power consumption at ultimate pressure 200 W 200 W 200 W
Admissible ambient temperature 10 to 55 °C / 50 to 131 °F 10 to 55 °C / 50 to 131 °F 10 to 55 °C / 50 to 131 °F
Cooling standard Water Water Water
Cooling optional Air Air Air
Cooling water connection 10 mm hose nozzle 10 mm hose nozzle 10 mm hose nozzle
Cooling water consumption 20 to 80 l/hr 20 to 80 l/hr 20 to 80 l/hr
Permissible cooling water pressure 3 to 7 bar 3 to 7 bar 3 to 7 bar
Permissible cooling water temperature 10 to 30 °C (50 to 86 °F) 10 to 30 °C (50 to 86 °F) 10 to 30 °C (50 to 86 °F)
Weight 25 kg (55 lbs)  25 kg (55 lbs)  25 kg (55 lbs)

Leybold TURBOVAC 1000C Dimensional Drawing:

Leybold TMP 1000C Pumping Speed:

Manuals:

Leybold Vacuum TMP 1000 Turbopump Manual

Leybold Vacuum TURBOVAC 1000 C Technical Data and Ordering Information

Leybold Vacuum TURBOVAC Classic Line Product Information

Leybold Vacuum TURBOVAC Line Operating Instructions